347
Lower Extremities
ExtrCombi
Indications:
Spiral mode for the combination of HiRes bone and
soft tissue studies, e.g. masses, trauma, disorders of
the joint etc.
A range of 15 cm will be done in 17.0 sec.
ExtrCombi 2
nd
reconstr.
kV 120
Effective mAs 90
Slice collimation 0.75 mm
Slice width 5.0 mm 1.0 mm
Feed/Rotation 6.6 mm
Rotation time 0.75 sec.
Kernel B60s B60s
Increment 5.0 mm 0.7 mm
CTDI
Vol
7.02 mGy
Effective dose Male: 0.01 mSv
Female: 0.01 mSv
For the 2
nd
reconstruction the Autoload into MPRthick
Range on the 3D Card is activated. The images will be
automatically loaded into 3D, MPRthick, and a coronal
MPRthick Range will pop up.
Please notice, if you are not satisfied with the Range
preset, adapt the parameters to your needs and link
them to the series.