Page view 34
Technical specifications 35
CEP08027: March 2009
GE LightSpeed
VCT
GE LightSpeed
VCT XT
Philips
Brilliance
CT 64
Siemens
SOMATOM
Sensation 64
Siemens
SOMATOM
Definition AS 64
Siemens
SOMATOM
Definition
Dual Source
Toshiba
Aquilion 64
13.13
Scanner settings
for std body scan,
used for quoting
CTDI100 values
120 kVp, 40
mm; 260 mAs;
0.4 to 1.0 sec
scan
64 x 0.625 or 32
x 0.625;
5 mm nominal
image thick
121 kVp, 40
mm; 260 mAs;
0.4 to 1.0 sec
scan
64 x 0.625 or 32
x 0.625;
5 mm nominal
image thick
120 kV, 250
mAs,
16x2.5mm,
5mm ST, 350
cm FOV, B
Filter
120 kV, 440
mA, 0.5 s, 4.8
mm slice (coll.
24x1.2 mm),
Large Focal
Spot
120 kV, 420 mA,
0.5 s, 4.8 mm
slice (coll. 16x1.2
mm), Standard
Shaped Beam
Filter
120 kV, 160
mA, 1.0 s, 6 mm
slice (coll. 6x3
mm), Standard
Shaped Beam
Filter
120 kV
300 mA
1 s/rot
16 mm
13.14
CTDI100 at
isocentre in air
[mGy / 100mAs]
Info not
available
Info. not
available
Info. not
available
18.1 21.3 20.7
Info not
available
13.15
CTDI100 at centre
of body phantom
[mGy/100mAs]
Small Body: 4.6
Medium Body:
5.1
Large Body: 4.8
Small Body: 4.6
Medium Body:
5.1
Large Body: 4.9
3.36 4.5 4.9 4.8
6.6 (Bow-tie
filter)
3.5 (Flat filter)
13.16
CTDI100 periphery
of body phantom
[mGy/100mAs]
Small Body: 8.7
Medium Body:
10.4
Large Body:
10.1
Small Body: 8.7
Medium Body:
10.4
Large Body:
10.1
6.76 7.6 8.8 8.5
14.8 (Bow-tie
filter)
7.3 (Flat filter)