Technical specifications 31
CEP08027: March 2009
GE LightSpeed
VCT
GE LightSpeed
VCT XT
Philips
Brilliance
CT 64
Siemens
SOMATOM
Sensation 64
Siemens
SOMATOM
Definition AS 64
Siemens
SOMATOM
Definition
Dual Source
Toshiba
Aquilion 64
12.7
Detector calibration
at warm-up
Air cal
recommended
Air cal
recommended
Yes, air cals Part of warm up Part of warm up Part of warm up Automatic
12.8
Time to perform
detector
calibrations [mins]
< 30 < 30 25 7 6 7 2
12.9
Recommended
frequency for
detector calibration
Once every 24
hours
Once every 24
hours
Weekly Daily Daily Daily Daily
12.10
Automated QC
tests
No No
Std QA
Protocols
Yes Yes Yes No
12.11
Self-check
frequency
Not applicable Not applicable Daily Daily Daily Daily Not applicable
12.12
Any additional
calibration by the
radiographer
No No
Info. not
provided
Not required Not required Not required Not required
12.13
Time to perform
these additional
calibrations [mins]
Not applicable Not applicable
Info. not
provided
Not required Not required Not required Not applicable