Page view 9
Technical specifications 10
CEP08027: March 2009
GE LightSpeed
VCT
GE LightSpeed
VCT XT
Philips
Brilliance
CT 64
Siemens
SOMATOM
Sensation 64
Siemens
SOMATOM
Definition AS
64
Siemens
SOMATOM
Definition
Dual Source
Toshiba
Aquilion 64
3.7
Half value layer of
x-ray beam at 120
kV (or nearest) [mm
Al]
7.5 [body] 6.3
[head]
7.5 [body] 6.3
[head]
6.9 ± 0.5 @ 120
KV
9.1 8.7 8.7
5.2 (L wedge
filter)
6.2 (M wedge
filter)
8.8 (DR wedge
filter)
3.8
Anode heat capacity
[MHU]
8 8 8 30 equivalent 30 equivalent 30 equivalent 7.5
3.9
Maximum anode
cooling rate
[kHU / min]
2100 2100 1608 5000 7300 7300 1386
3.10
Method of tube
cooling
Oil to air Oil to air Oil to air Direct oil Direct oil Direct oil Liquid
3.11
Expected tube life
[amp-hours]
6,000 exams 6,000 exams
1 year unlimited
guarantee
3 years (see
note 1)
3 years (see
note 1)
3 years (see
note 1)
300000
rotations
Table 4. Detection system
GE LightSpeed
VCT
GE LightSpeed
VCT XT
Philips
Brilliance
CT 64
Siemens
SOMATOM
Sensation 64
Siemens
SOMATOM
Definition AS
64
Siemens
SOMATOM
Definition
Dual Source
Toshiba
Aquilion 64
4.1 Detector type Solid state array Solid state array Solid state array Solid state array
Solid state
array
Solid state array Solid state array
Page view 9