Siemens CT User Manual Page 34

  • Download
  • Add to my manuals
  • Print
  • Page
    / 98
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 33
Technical specifications 34
CEP08027: March 2009
GE LightSpeed
VCT
GE LightSpeed
VCT XT
Philips
Brilliance
CT 64
Siemens
SOMATOM
Sensation 64
Siemens
SOMATOM
Definition AS 64
Siemens
SOMATOM
Definition
Dual Source
Toshiba
Aquilion 64
13.9
Scanner settings
for std head scan,
used for quoting
CTDI100 values
120 kVp, 40
mm; 260 mAs;
0.4 to 1.0 sec
scan
64 x 0.625 or 32
x 0.625; 5 mm
nominal image
thick
120 kVp, 40
mm; 260 mAs;
0.4 to 1.0 sec
scan
64 x 0.625 or 32
x 0.625; 5 mm
nominal image
thick
120 kV, 200
mAs,
16x1.5mm,
6mm ST, 250
cm FOV, EB
Filter
120 kV, 330
mA, 1.0s, 4.8
mm slice (coll.
24x1.2 mm),
Small Focal
Spot
120 kV, 350 mA,
1.0 s, 4.8 mm
slice (coll. 12x1.2
mm), Standard
Shaped Beam
Filter
120 kV, 400
mA, 1.0 s, 6 mm
slice (coll. 6x3
mm), Narrow
Shaped Beam
Filter
120 kV
300 mA
1 s/rot
16 mm
13.10
CTDI100
at
isocentre in air
[mGy / 100mAs]
Info not
available
Info not
available
Info. not
available
18.1 22.3 19.4
Info not
available
13.11
CTDI100
at centre
of head phantom
[mGy / 100mAs]
Small Head:
16.9:
Head: 17.8
Small Head:
16.9:
Head: 17.8
10.36 13.4 16.4 14
18.4 (Bow-tie
filter)
10.0 (Flat filter)
13.12
CTDI100 at
periphery of head
phantom [mGy /
100mAs]
Small Head:
17.0
Head: 20.0
Small Head:
17.0
Head: 20.0
11.24 13.6 16.7 13.3
21.5 (Bow-tie
filter)
11.1 (Flat filter)