2-67
1010NFM-3J
Section 2
2.11.3 THE APPLICATION INFO MENU
This menu provides a live display of the basic timing data used by meter during operation.
2.11.4 THE LIQUID DATA MENU
This menu shows the current Reynolds number used by the meter to implement flow profile compen-
sation. Depending on the 1010 model, Temp 1 & 2, Viscosity cP, Viscosity cS and Pressure are also
offered.
TN uSec 75.193
TL uSec 44.009
DeltaT uSec 126
Burst/Sec 256
% Accepted 100
Last Makeup (Re)Start
Makeup Status Measurement
Siemens 2 Channel [1] 1
Time between transit and receive
Application Info
Temp 1 Current Liquid Temperature
Temp 2 Current Liquid Temperature
Reynolds # Current Reynolds number
Specific Gravity Current specific gravity
Viscosity cP The absolute viscosity of a liquid dynamically measured in centipoise.
Pressure Derived spanned from analog input.
Viscosity cS Kinematic viscosity of a liquid measured in centistokes.
LIQUID DATA MENU ITEMS
APPLICATION INFO MENU ITEMS
TN uSec The total elapsed time between the transmission and reception of a sonic pulse.
TL uSec, The total time a sonic pulse takes to travel through the liquid.
DeltaT nSec This is the instantaneous result of subtracting the down from the up transit times.
NOTE: This will appear to be more active than dampened flow readings.
Burst/Sec Number of transmissions per second under current operating conditions.
The following factors influence this parameter: aeration, pressure transients,
signal strength variation.
% Accepted The running tally of accepted and rejected up/down transmit burst sets.
Last Makeup Reason for last Makeup (signal reacquisition).
Makeup Status Current stage of Makeup routine. [Measurement = normal operation]