2-61
1010NFM-3J
Section 2
To assign functions to Relay 1:
To access the [Relay] option list press
To move the cursor to the desired Relay assignment press
To store selection press . Repeat procedure for Relay 2.
ENT
2.10.3 ANALOG INPUT SETUP (optional function)
The Analog Input Setup function assigns an active analog input to a measurement channel/path. The
meter provides four DC current input ports for single and Dual Channel units. The DC current input
ranges from a zero level of 4 mA to a full scale of 20 mA. The [Analog Input Setup] menu cell allows you
to enable this port and then span it to any desired scaling.
For example, when using the analog input viscosity function the numeric variables might be spanned
as follows: 4mA=1 (water) and 20mA=100 (more viscous liquid). This spanning configuration allows
the meter to use this constant numerical change to improve calibration in real time. The various flow-
meter models allow you to associate the analog input to active system variables such as specific
gravity, viscosity and others (see table on next page).
NOTE: Using the <F1> key (Totalizer clear command) also clears all channel totalizers plus
the overflow register described in the last paragraph.
Assigning Relay 1 and 2 Functions
System 1010N, depending on the model, provides at least two alarm relays. Please refer to the Hard-
ware Installation Drawing for wiring details. Relays respond to any of the alarm conditions/data func-
tions included on the Relay Option List.
Not Used Not Active
Power Off Power Off alarm occurs when power fails
S.G. Specific Gravity value relay trip-point
Hi Temperature High temperature value relay trip-point
Low Temperature Low temperature value relay trip-point
High Flow Flow rate exceeds high flow setpoint
Low Flow Flow rate falls below low flow setpoint
Flow Alarm Flow rate exceeds or falls below flow setpoints
Fault Alarm System loses receive signal
Spacing Transducer spacing needs adjusting
Empty Empty pipe alarm
Aeration Aeration percentage exceeds alarm setpoint
Reverse Flow Flow is in negative direction
BatchTot Batch/Sample total advances
Interface Liquid interface setpoint exceeded
Pos Total Positive total volume advances by 1 digit
Neg Total Negative total volume advances by 1 digit
Soft Fault Fault condition - memory mode active
RELAY OPTION LIST