HeartView CT
How to do it
Calcium Scoring
This application is used for identification and quan-
tification of calcified lesions in the coronary arteries.
It can be performed with both Prospective ECG trigger-
ing (sequen-tial scanning) and Retrospective gating
(spiral scanning) techniques. The following scan pro-
tocols are predefined:
CaScoreRoutine
Standard spiral protocol with ECG-gating which
has been clinically used so far, relying on 12-slice
acquisition, 1.5 mm collimation and 0.42 s gantry
rotation time. The table feed is 5.6 mm/rotation.
CaScoreFastVolume
Protocol with ECG-gating using 16-slice acquisition,
1.5 mm collimation and 0.42 s gantry rotation time.
With the superior ability of the 16-slice acquisition,
the table feed is increased to 6.8 mm/rotation.
The time to cover a typical scan range of 15 cm will
therefore be reduced to less than 10 s.
CaScoreSeq
Sequential scan protocol with ECG-triggering and
simultaneous acquisition of six 3 mm slices.
Page view 37