Siemens SOMATOM Sensation Cardiac Version A60 Operations Instructions Page 20

  • Download
  • Add to my manuals
  • Print
  • Page
    / 142
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 19
20
General
Workflow
Fig. 2: Principle of CARE Dose tube current adaptation.
lateral
1.00
0.75
0.50
0.25
rel. units
a.p.
time
Object attenuation
Modulated tube current
CARE Dose is pre-selected by default for most standard
protocols, except head and cardiac protocols. It can
be switched on/off in the scan card.
The application of CARE Dose does not require any
changes in the scan parameters. The mAs, however,
should be adapted to patient size manually for obese
and pediatric patients.
The mean value of the mAs applied will be lower than
what you have selected. Although the average mA
for the entire scan will be lower than selected, we
allow the scanner to apply increased mA levels for the
high attenuation views. This may cause different
results of the tube load controller when switching on
and off CARE Dose.
The mean value of the effective mAs applied is shown
in the image text.
Page view 19
1 2 ... 15 16 17 18 19 20 21 22 23 24 25 ... 141 142

Comments to this Manuals

No comments